Our NEW TIGER OPTICS™ HALO Max QCL™ allows dramatic decreases in detection limits for certain molecules, such as CO. The HALO Max QCL CO for ppt-level carbon monoxide detection is based on our latest Max platform, offers exceptional speed and further improved usability in an all-inclusive and robust package. The analyzer is fast to install, offers continuous, real-time detection, and is easy to use and effortless to maintain, with built-in zero verification and zero drift. The HALO Max QCL CO perfectly complement our HALO KA Max series (for H2O, NH3 and CH4), the HALO OK (for O2) and our other QCL-based analyzer, the HALO Max QCL CO2 to utilize the advantages of CRDS for detection of a large variety of critical trace impurities.
Detection and Matrix | Range | LDL (3σ) | Precision (1σ) @ zero |
---|---|---|---|
CO in N2 | 0 – 0.5 ppm | 200 ppt | 70 ppt |
CO in He | 0 – 0.35 ppm | 130 ppt | 45 ppt |
CO in Ar | 0 – 0.4 ppm | 150 ppt | 50 ppt |
CO in H2 | 0 – 0.5 ppm | 200 ppt | 70 ppt |
CO in O2 | 0 – 0.45 ppm | 170 ppt | 60 ppt |
CO in Clean Dry Air (CDA) | 0 – 0.5 ppm | 200 ppt | 70 ppt |
Performance | |
---|---|
Operating range | See Detection Capability table |
Detection Limit (LDL) | See Detection Capability table |
Precision (1σ, greater of) | ± 0.75% or 1/3 of LDL |
Accuracy (greater of) | ± 4% or LDL |
Speed of response | < 1 minute to 95% |
Environmental conditions | 10°C to 40°C, 30% to 80% RH (non-condensing) |
Storage temperature | -10°C to 50°C |
Gas Handling System and Conditions | |
Gas connections | 1/4” male VCR inlet and outlet |
Leak tested to | 1 x 10-9 mbar l / sec |
Inlet pressure | 6 − 125 psig (1.4 − 9.6 bara) |
Flow rate | <1 slpm in N2 (gas dependent) |
Sample gases | Most inert and passive gases |
Gas temperature | Up to 60°C |
Dimensions, H x W x D | |
Standard sensor | 8.75″ x 19.0″ x 24.0″ (222 mm x 483 mm x 610 mm) |
Weight | |
Standard sensor | 55 lbs (25 kg) |
Electrical and Interfaces | |
Platform | Max series analyzer |
Alarm indicators | 2 user programmable, 1 system fault, Form C relays |
Power requirements | 90 − 240 VAC, 50/60 Hz |
Power consumption | 100 Watts max. |
Signal output | Isolated 4−20 mA |
User interfaces | 5.7” LCD touchscreen, 10/100 Base-T Ethernet |
USB, RS-232, RS-485, Modbus TCP (optional) | |
Data storage | Internal or external flash drive |
Certification | CE Mark |
Patents | |
U.S. Patent #7,277,177 |
Watch this video and see how fast it really is.
Learn about our service, technical support, spares, and our CRDS Analyzer | Performance Verification Program
Analyzer in service? Need a rental CRDS analyzer? Ask us how we can help.
Americas +1 (215) 656-4000
EMEA +49 69 20436910
China/APAC +86 400 086 0106
Copyright © 2023 Process Insights, Inc. All Rights Reserved.