HALO 3 HCI
Our TIGER OPTICS™ HALO™ 3 HCl trace level hydrogen chloride gas analyzer provides users with the unmatched accuracy, reliability, speed of response and ease of operation. Featuring our proven Cavity Ring-Down Spectroscopy-based trace gas sensor in a very compact and economic analyzer design, this versatile instrument allows users to measure HCl in most inert and passive gases with just one device. Users also enjoy freedom from requirements, such as periodic sensor maintenance, span calibrations, purifier replacement and pump rebuild. As a result, the HALO is ideally suited to many applications where HCl impurities are extremely critical, such as semiconductor utilization.
Detection and Matrix | Range | LDL (3σ) | Precision (1σ) @ zero |
---|---|---|---|
HCl in N2 | 0 – 20 ppm | 1.0 ppb | 0.4 ppb |
HCl in Clean Dry Air (CDA) | 0 – 20 ppm | 1.0 ppb | 0.4 ppb |
HCl in H2 | 0 – 10 ppm | 1.0 ppb | 0.4 ppb |
Performance | |
---|---|
Operating range | See Detection Capability table |
Detection limit (LDL, 3σ/24h) | See Detection Capability table |
Precision (1σ, greater of) | ± 0.75% or 1/3 of LDL |
Accuracy (greater of) | ± 4% or LDL |
Speed of response | < 1 minute to 90% |
Environmental conditions | 10°C to 40°C, 30% to 80% RH (non-condensing) |
Storage temperature | -10°C to 50°C |
Gas Handling System and Conditions | |
Wetted materials | 316L stainless steel (optional corrosion-resistant material), 10 Ra surface finish |
Gas connections | 1/4” male VCR inlet and outlet |
Leak tested to | 1 x 10-9 mbar l / sec |
Inlet pressure | 10 − 125 psig (1.7 − 9.6 bara) |
Flow rate | Up to 1.8 slpm |
Sample gases | Most inert, toxic, passive and corrosive matrices |
Gas temperature | Up to 60°C |
Dimensions, H x W x D | |
Standard sensor | 8.73″ x 8.57″ x 23.6″ (222 mm x 218 mm x 599 mm) |
Sensor rack (fits up to two sensors) | 8.73″ x 19.0″ x 23.6″ (222 mm x 483 mm x 599 mm) |
Weight | |
Standard sensor | 28 lbs (12.7 kg) |
Electrical and Interfaces | |
Platform | Max series analyzer |
Alarm indicators | 2 user programmable, 1 system fault, Form C relays |
Power requirements | 90 − 240 VAC, 50/60 Hz |
Power consumption | 40 Watts max. |
Signal output | Isolated 4−20 mA per sensor |
User interfaces | 5.7” LCD touchscreen, 10/100 Base-T Ethernet |
UBS, RS-232, RS-485, Modbus TCP (optional) | |
Data storage | Internal or external flash drive |
Certification | CE Mark |
Patents | |
U.S. Patent #7,277,177 |
Serani™ Analyzer Interface Software offers users a new way to access their analyzer remotely from any computer or laptop. Compatible with most HALO, Spark, Tiger-i, CO-rekt and ALOHA H2O series analyzers.
Americas +1 (215) 656-4000
EMEA +49 69 20436910
China/APAC +86 400 086 0106
Copyright © 2022 Process Insights, Inc. All Rights Reserved.