HALO 3 NH3
Ammonia (NH3) is a key impurity in many applications, ranging from industrial process control to the analysis of fuel cell hydrogen. Our TIGER OPTICS™ HALO™ 3 NH3 gas analyzer delivers low detection limits, drift-free operation, and compatibility with many different sample gases make it an ideal tool for monitoring trace amounts of ammonia, for example, to ensure compliance with SAE J2719, ISO 14687 or similar purity standards for hydrogen used for fuel cell electric vehicles (FCEVs).
Highly specific to the target molecule, CW-CRDS also prevents cross-interferences from distorting your measurement. Plus, there is no need to perform costly and time-consuming zero and span calibrations, saving both time and money with continuous, online service.
Detection and Matrix | Range | LDL (3σ) | Precision (1σ) @ zero |
---|---|---|---|
Standard Model | |||
NH3 in N2 (Low Range) | 0 – 7 ppm | 0.5 ppb | 0.2 ppb |
NH3 in N2 (Mid-Range) | 0 – 35 ppm | 2.5 ppb | 0.8 ppb |
NH3 in N2 (High Range) | 0 – 130 ppm | 20 ppb | 7 ppb |
NH3 in H2 (Low Range) | 0 – 6 ppm | 0.4 ppb | 0.15 ppb |
NH3 in H2 (Mid-Range) | 0 – 30 ppm | 2.0 ppb | 0.7 ppb |
NH3 in H2 (High Range) | 0 – 110 ppm | 15 ppb | 5 ppb |
NH3 in CO2† | 0 – 30 ppm | 2.5 ppb | 0.8 ppb |
N2O Model | |||
NH3 in N2 | 0 – 150 ppm | 9 ppb | 3 ppb |
NH3 in N2O | 0 – 200 ppm | 10 ppb‡ / 50 ppb | 3.5 ppb‡ / 20 ppb |
†Cannot be combined with Low or High Range detection in N2/H2
‡Dry vacuum pump required
Performance | |
---|---|
Operating range | See Detection Capability table |
Detection limit (LDL, 3σ/24h) | See Detection Capability table |
Precision (1σ, greater of) | ± 0.75% or 1/3 of LDL |
Accuracy (greater of) | ± 4% or LDL |
Speed of response | < 3 minutes to 95% |
Environmental conditions | 10°C to 40°C, 30% to 80% RH (non-condensing) |
Storage temperature | -10°C to 50°C |
Gas Handling System and Conditions* | |
Wetted materials | 316L stainless steel, 10 Ra surface finish |
Gas connections | 1/4″ male VCR inlet and outlet |
Leak tested to | 1 x 10-9 mbar l / sec |
Inlet pressure | 10 – 125 psig (1.7 – 9.6 bara) |
Flow rate | ∼1 slpm |
Sample gases | Most inert, toxic, passive and corrosive matrices |
Gas temperature | Up to 60°C |
Dimensions, H x W x D | |
Standard sensor | 8.73″ x 8.57″ x 23.6″ (222 mm x 218 mm x 599 mm) |
Sensor rack (fits up to two sensors) | 8.73″ x 19.0″ x 23.6″ (222 mm x 483 mm x 599 mm) |
Weight | |
Standard sensor | 34 lbs (15.4 kg) |
Electrical and Interfaces | |
Platform | Max series analyzer |
Alarm indicators | 2 user programmable, 1 system fault, Form C relays |
Power requirements | 90 − 240 VAC, 50/60 Hz |
Power consumption | 40 Watts max. |
Signal output | Isolated 4−20 mA per sensor |
User interfaces | 5.7” LCD touchscreen, 10/100 Base-T Ethernet |
USB, RS-232, RS-485, Modbus TCP (optional) | |
Data storage | Internal or external flash drive |
Certification | CE Mark |
Patents | |
U.S. Patent #7,277,177 |
*Analysis in some specialty gases and certain applications may require a vacuum pump for operation. Please contact us to discuss your specific requirements.