HALO H2O in N2O
Our TIGER OPTICS™ HALO™ H2O in N2O analyzer provides users with the unmatched accuracy, reliability, speed of response and ease of operation. Featuring our powerful Cavity Ring-Down Spectroscopy-based moisture sensor in a compact and affordable package, this versatile analyzer allows users to measure trace moisture in nitrous oxide (N2O) and other gases.
Users can also expect significant cost savings, with no need for periodic sensor maintenance, span calibrations, purifier replacements, and pump rebuilds. As a result, the HALO H2O in N2O analyzer is ideally suited to many applications where moisture measurement is extremely critical. These applications include process control, continuous quality control, and cylinder analysis for medical and calibration gases and more.
Detection and Matrix | Range | LDL (3σ) | Precision (1σ) @ zero |
---|---|---|---|
H2O in N2 | 0 – 15 ppm | 2.2 ppb | 0.8 ppb |
H2O in N2O | 0 – 20 ppm | 7.5 ppb | 2.5 ppb |
H2O in Ar | 0 – 6 ppm | 1.0 ppb | 0.35 ppb |
H2O in He | 0 – 3 ppm | 0.5 ppb | 0.2 ppb |
H2O in H2 | 0 – 12 ppm | 1.9 ppb | 0.7 ppb |
H2O in O2 | 0 – 8 ppm | 1.2 ppb | 0.4 ppb |
Performance | |
---|---|
Operating range | See Detection Capability table |
Detection limit (LDL, 3σ/24h) | See Detection Capability table |
Precision (1σ, greater of) | ± 0.75% or 1/3 of LDL |
Accuracy (greater of) | ± 4% or LDL |
Speed of response | < 1 minute to 90% |
Environmental conditions | 10°C to 40°C, 30% to 80% RH (non-condensing) |
Storage temperature | -10°C to 50°C |
Gas Handling System and Conditions | |
Wetted materials | 316L stainless steel, 10 Ra surface finish |
Gas connections | 1/4″ male VCR inlet and outlet |
Leak tested to | 1 x 10-9 mbar l / sec |
Inlet pressure | 10 – 125 psig (1.7 – 9.6 bar) |
Flow rate | Up to 1.8 slpm |
Sample gases | Most inert, passive, toxic and corrosive gases |
Gas temperature | Up to 60°C |
Dimensions, H x W x D | |
Standard sensor | 8.73″ x 8.57″ x 23.6″ (222 mm x 218 mm x 599 mm) |
Sensor rack (fits up to two sensors) | 8.73″ x 19.0″ x 23.6″ (222 mm x 483 mm x 599 mm) |
Standard sensor | 28 lbs (12.7 kg) |
Electrical and Interfaces | |
Platform | Max series analyzer |
Alarm indicators | 2 user programmable, 1 system fault, Form C relays |
Power requirements | 90 – 240 VDC, 50/60 Hz |
Power consumption | 40 Watts max. |
Signal output | Isolated 4–20 mA per sensor |
User interfaces | 5.7″ LCD touchscreen, 10/100 Base-T Ethernet |
USB, RS-232, RS-485, Modbus TCP (optional) | |
Data storage | Internal or external flash drive |
Certification | CE Mark |
Patents | |
U.S. Patent #7,277,177 |
Serani™ Analyzer Interface Software offers users a new way to access their analyzer remotely from any computer or laptop. Compatible with most HALO, Spark, Tiger-i, CO-rekt and ALOHA H2O series analyzers.
Americas +1 (215) 656-4000
EMEA +49 69 20436910
China/APAC +86 400 086 0106
Copyright © 2022 Process Insights, Inc. All Rights Reserved.