For the first time, powerful advanced spectroscopy is available at a popular price for a host of applications, from quality assurance to cylinder filling, as well as welding, medical, industrial and high-purity gas production; bulk delivery and distribution transfer points; and more. Say goodbye to cumbersome, complex, costly and labor-intensive mid-20th century technology. Gone is the need for calibration, spare parts, cramped ranges, and worries about drift and downtime. The speed alone will make you gasp. Plus, it’s a joy to start-up and to operate.
Detection and Matrix | Range | LDL (3σ) | Precision (1σ) @ zero |
---|---|---|---|
H2O in N2 | 0 – 2000 ppm | 12 ppb | 4 ppb |
H2O in O2 | 0 – 1000 ppm | 6 ppb | 2 ppb |
H2O in Ar | 0 – 900 ppm | 4.5 ppb | 1.5 ppb |
H2O in He | 0 – 450 ppm | 3 ppb | 1.0 ppb |
H2O in H2 | 0 – 1750 ppm | 7.5 ppb | 2.5 ppb |
H2O in Clean Dry Air (CDA) | 0 – 1800 ppm | 10 ppb | 3 ppb |
H2O in Ne | 0 – 450 ppm | 30 ppb | 10 ppb |
H2O in Kr | 0 – 1100 ppm | 5.5 ppb | 1.8 ppb |
H2O in Xe | 0 – 1300 ppm | 7.5 ppb | 2.5 ppb |
H2O in CO | 0 – 2400 ppm | 15 ppb | 5 ppb |
H2O in CF4 | 0 – 1300 ppm | 9 ppb | 3 ppb |
H2O in NF3 | 0 – 1800 ppm | 9 ppb | 3 ppb |
H2O in SF6 | 0 – 1300 ppm | 15 ppb | 5 ppb |
Performance | |
---|---|
Operating range | See Detection Capability table |
Detection Limit (LDL, 3σ/24h) | See Detection Capability table |
Precision (1σ, greater of) | ± 0.75% or 1/3 of LDL |
Accuracy (greater of) | ± 4% or LDL |
Speed of response | < 3 minutes to 90% |
Environmental conditions | 10°C to 40°C, 30% to 80% RH (non-condensing) |
Storage temperature | -10°C to 50°C |
Gas Handling System and Conditions | |
Wetted materials | 316L stainless steel, 10 Ra surface finish |
Gas connections | 1/4″ male VCR inlet and outlet |
Inlet pressure* | 10 – 125 psig (1.7 – 9.6 bar) |
Flow rate | ∼1.0 slpm (in N2) |
Sample gases | Most inert, toxic, and passive matrices |
Gas temperature | Up to 60°C |
Dimensions, H x W x D | |
Standard sensor | 8.73″ x 8.57″ x 23.6″ (222 mm x 218 mm x 599 mm) |
Sensor rack (fits up to two sensors) | 8.73″ x 19.0″ x 23.6″ (222 mm x 483 mm x 599 mm) |
Weight | |
Standard sensor | 32 lbs (14.5 kg) |
Electrical and Interfaces | |
Platform | Max series analyzer |
Alarm indicators | 2 user programmable, 1 system fault, Form C relays |
Power requirements | 90 – 240 VAC, 50/60 Hz |
Power consumption | 40 Watts max. |
Signal output | Isolated 4–20 mA per sensor |
User interfaces | 5.7″ LCD touchscreen, 10/100 Base-T Ethernet |
USB, RS-232, RS-485, Modbus TCP (optional) | |
Data storage | Internal and external flash drive |
Certification | CE Mark |
Patents | |
U.S. Patent #7,277,177 | |
*See brochure for lower-pressure options |