HALO H2O in N2O
Trace Moisture CRDS Analyzer in Pure Nitrous Oxide
CRDS Laser-Based Analyzer
PPB-LEVEL MOISTURE ANALYZER FOR PURE NITROUS OXIDE
Our TIGER OPTICS™ HALO™ H2O in N2O analyzer provides users with the unmatched accuracy, reliability, speed of response and ease of operation. Featuring our powerful Cavity Ring-Down Spectroscopy-based moisture sensor in a compact and affordable package, this versatile analyzer allows users to measure trace moisture in nitrous oxide (N2O) and other gases.
Users can also expect significant cost savings, with no need for periodic sensor maintenance, span calibrations, purifier replacements, and pump rebuilds. As a result, the HALO H2O in N2O analyzer is ideally suited to many applications where moisture measurement is extremely critical. These applications include process control, continuous quality control, and cylinder analysis for medical and calibration gases and more.
DESIGNED FOR TRACE LEVEL MOISTURE ANALYSIS IN NITROUS OXIDE
- Low parts per billion (ppb) moisture detection capability in medical N2O
- Absolute measurement (freedom from calibration gases)
- Wide dynamic range—over four orders of magnitude
- Low cost of ownership and operational simplicity
- Clean technology—no external calibration gases required
- Compact analyzer footprint
APPLICATIONS
- Gas Quality Control
- Research & Development
- Measure trace moisture in nitrous oxide and other gases
- Process and quality control, and cylinder analysis for medical and calibration gases
DETECTION CAPABILITY
Detection and Matrix | Range | LDL (3σ) | Precision (1σ) @ zero |
---|---|---|---|
H2O in N2 | 0 – 15 ppm | 2.2 ppb | 0.8 ppb |
H2O in N2O | 0 – 20 ppm | 7.5 ppb | 2.5 ppb |
H2O in Ar | 0 – 6 ppm | 1.0 ppb | 0.35 ppb |
H2O in He | 0 – 3 ppm | 0.5 ppb | 0.2 ppb |
H2O in H2 | 0 – 12 ppm | 1.9 ppb | 0.7 ppb |
H2O in O2 | 0 – 8 ppm | 1.2 ppb | 0.4 ppb |
SPECIFICATIONS
Performance | |
---|---|
Operating range | See Detection Capability table |
Detection limit (LDL, 3σ/24h) | See Detection Capability table |
Precision (1σ, greater of) | ± 0.75% or 1/3 of LDL |
Accuracy (greater of) | ± 4% or LDL |
Speed of response | < 1 minute to 90% |
Environmental conditions | 10°C to 40°C, 30% to 80% RH (non-condensing) |
Storage temperature | -10°C to 50°C |
Gas Handling System and Conditions | |
Wetted materials | 316L stainless steel, 10 Ra surface finish |
Gas connections | 1/4″ male VCR inlet and outlet |
Leak tested to | 1 x 10-9 mbar l / sec |
Inlet pressure | 10 – 125 psig (1.7 – 9.6 bar) |
Flow rate | Up to 1.8 slpm |
Sample gases | Most inert, passive, toxic and corrosive gases |
Gas temperature | Up to 60°C |
Dimensions, H x W x D | |
Standard sensor | 8.73″ x 8.57″ x 23.6″ (222 mm x 218 mm x 599 mm) |
Sensor rack (fits up to two sensors) | 8.73″ x 19.0″ x 23.6″ (222 mm x 483 mm x 599 mm) |
Standard sensor | 28 lbs (12.7 kg) |
Electrical and Interfaces | |
Platform | Max series analyzer |
Alarm indicators | 2 user programmable, 1 system fault, Form C relays |
Power requirements | 90 – 240 VDC, 50/60 Hz |
Power consumption | 40 Watts max. |
Signal output | Isolated 4–20 mA per sensor |
User interfaces | 5.7″ LCD touchscreen, 10/100 Base-T Ethernet |
USB, RS-232, RS-485, Modbus TCP (optional) | |
Data storage | Internal or external flash drive |
Certification | CE Mark |
Patents | |
U.S. Patent #7,277,177 |
Learn about our service, technical support, spares, and our CRDS Analyzer | Performance Verification Program
Analyzer in service? Need a rental CRDS analyzer? Ask us how we can help.
ANALYZE WITH EASE™
TRACE MOISTURE ANALYZER FOR PURE NITROUS OXIDE
ANALYZE WITH EASE™
TRACE MOISTURE ANALYZER FOR PURE NITROUS OXIDE
SALES | TRAINING INQUIRIES
AMERICAS: info.americas@process-insights.com
EMEAI (includes India): info.emeai@process-insights.com
APAC: info.apac@process-insights.com
CHINA: info.cn@process-insights.com