ADVANCING ACCURATE, CONSISTENT & DRIFT-FREE CO2 MEASUREMENTS
The removal of contaminants prior to cooling and distillation is essential to the cryogenic air separation process. If not detected quickly, impurities such as CO2 (carbon dioxide) can freeze in the downstream cryogenic equipment causing damage and product spoilage. Our TIGER OPTICS™ HALO™ 3 CO2 analyzer affords fast, accurate response and clean-up, with no possibility of drift. Based on powerful Cavity Ring-Down Spectroscopy (CRDS), with a proprietary laser-locked cell, the HALO 3 is free of drift, guaranteeing consistent and reliable trace CO2 detection in nitrogen and other inert gases. Highly specific to the target molecule, CRDS also prevents cross interferences from distorting your measurement. Plus, there is no need to perform costly and time-consuming zero and span calibrations, saving both time and money with continuous, on-line service.
Based on powerful Cavity Ring-Down Spectroscopy (CRDS), with a proprietary laser-locked cell, the HALO 3 is free of drift, guaranteeing consistent and reliable trace CO2 detection in nitrogen and other inert gases. Highly specific to the target molecule, CRDS also prevents cross interferences from distorting your measurement. Plus, there is no need to perform costly and time-consuming zero and span calibrations, saving both time and money with continuous, on-line service.
Compact and portable, the HALO 3 CO2 gives you unsurpassed speed of response and ease of use. In sum, the Halo 3 CO2 analyzer serves a range of applications where trace gas measurement is extremely critical, such as syngas production, fixed bulk gas continuous monitoring, gas cylinder quality control, auto-load truck fill and a multitude of other challenging applications. The HALO 3 CO2 builds on our longstanding leadership for trace monitoring of critical compounds in pressurized gases.
BEST-IN-CLASS PERFORMANCE ANND RELIABILITY
- Lower detection limit down to 8 ppb in nitrogen
- Wide dynamic range
- Freedom from drift
- No spectral interference
- Compact standalone footprint or rack mountable
- Low Cost of Ownership
- Simple operation
APPLICATIONS
- Air Separation Units & Nitrogen Plants
- Gaseous CRM & Calibration Gas Mixtures
- High-Purity Gases & Systems
- Hydrogen Fuel Cells
- Research & Development
- Syngas & Fuel Gas Production
DETECTION CAPABILITY
Detection and Matrix | Range | LDL (3σ) | Precision (1σ) @ zero |
---|---|---|---|
CO2 in N2 | 0 – 25 ppm | 8 ppb | 3 ppb |
CO2 in He | 0 – 25 ppm | 8 ppb | 3 ppb |
CO2 in Ar | 0 – 25 ppm | 8 ppb | 3 ppb |
CO2 in O2 | 0 – 25 ppm | 8 ppb | 3 ppb |
CO2 in Clean Dry Air (CDA) | 0 – 25 ppm | 8 ppb | 3 ppb |
CO2 in HCl | 0 – 30 ppm | 9 ppb | 3 ppb |
SPECIFICATIONS
Performance | |
---|---|
Operating range | See Detection Capability table |
Detection limit (LDL, 3σ/24h) | See Detection Capability table |
Precision (1σ, greater of) | ± 0.75% or 1/3 of LDL |
Accuracy (greater of) | ± 4% or LDL |
Speed of response | < 3 minute to 95% |
Environmental conditions | 10°C to 40°C, 30% to 80% RH (non-condensing) |
Storage temperature | -10°C to 50°C |
Gas Handling System and Conditions | |
Wetted materials | 316L stainless steel (optional corrosion-resistant material), 10 Ra surface finish |
Gas connections | 1/4” male VCR inlet and outlet |
Leak tested to | 1 x 10-9 mbar l / sec |
Inlet pressure | 10 − 125 psig (1.7 − 9.6 bara) |
Flow rate | Up to 1.8 slpm |
Sample gases | Most inert, toxic, passive and corrosive matrices |
Gas temperature | Up to 60°C |
Dimensions, H x W x D | |
Standard sensor | 8.73″ x 8.57″ x 23.6″ (222 mm x 218 mm x 599 mm) |
Sensor rack (fits up to two sensors) | 8.73″ x 19.0″ x 23.6″ (222 mm x 483 mm x 599 mm) |
Weight | |
Standard sensor | 28 lbs (12.7 kg) |
Electrical and Interfaces | |
Platform | Max series analyzer |
Alarm indicators | 2 user programmable, 1 system fault, Form C relays |
Power requirements | 90 − 240 VAC, 50/60 Hz |
Power consumption | 40 Watts max. |
Signal output | Isolated 4−20 mA per sensor |
User interfaces | 5.7” LCD touchscreen, 10/100 Base-T Ethernet |
USB, RS-232, RS-485, Modbus TCP (optional) | |
Data storage | Internal or external flash drive |
Certification | CE Mark |
Patents | |
U.S. Patent #7,277,177 |
Serani™ Max Analyzer Interface Software allows you to have remote access from your computer or laptop for our latest CRDS generation analyzers, such as the HALO KA Max and T-I Max. The Windows-based software package communicates with the analyzer through Ethernet and allows full control of the system conveniently from your desk.
ANALYZE WITH EASE™
LOW PPB CARBON DIOXIDE DETECTION
ANALYZE WITH EASE™
LOW PPB CARBON DIOXIDE DETECTION
Learn about our service, technical support, spares, and our CRDS Analyzer | Performance Verification Program
Analyzer in service? Need a rental CRDS analyzer? Ask us how we can help.
SALES | TRAINING INQUIRIES
AMERICAS: info.americas@process-insights.com
EMEAI (includes India): info.emeai@process-insights.com
APAC: info.apac@process-insights.com
CHINA: info.cn@process-insights.com
solutions flyers
-
AGRICULTURE
-
CHEMICAL
-
ENERGY
-
FOOD & BEVERAGE
-
FLARE GAS
-
INDUSTRIAL GAS
-
LAB & RESEARCH
-
PHARMACEUTICALS & LIFE SCIENCE
-
REFINERY
-
SEMICONDUCTOR
-
SPECIALTY GAS